Atomic Absorption Spectrophotometer

Used for testing the metal element concentration analyze in agriculture, chemical, environmental study, food, mining, and petrochemical, pharmaceutical industry.

Description

Atomic Absorption Spectrophotometer is an integrated flame/graphite furnace atomization system where the graphite system is fully automatic. Flame temperature is continuously adjustable between 2300-2900ºC, which makes it possible to choose the best atomization temperature. Rich Oxygen Flame will not pollute the environment and is not harmful for human bodies. It’s a break-through in flame AAS analysis.

Feature 

  • The changeover of the integrated flame and graphite furnace atomizer is automatically controlled featuring easy operation and time saving which eliminates human labor
  • To perform flame emission analysis, flame emission burner head can be installed for alkali metals as K, Na etc.
  • Provided with automatic 6-lamp turret, automatic adjustment of lamp current and optimization of light beam position
  • Parameters like Wavelength scanning, peak picking, change in spectral bandwidth, optimization of position parameters, automatic ignition and gas flow setting are done automatically
  • The graphite furnace analysis is fully reliable and automatic where it is adopting FUZZY-PID and dual curve mode light-controlled temperature control technique, temperature auto-correction technique, ensures fast heating, good temperature reproducibility and high analytical sensitivity
  • The temperature control accuracy is less than 1%
  • Graphite furnace with pneumatic control and pressure lock ensures constant pressure and reliable contact
  • Automatic standard sample preparation, automatic correction of sampling probe depth, automatic tracing and correction of liquid surface height in the sample vessel, with sampling accuracy of 1% and reproducibility of 0.3%
  • Equipped with alarm and automatic protection to fuel gas leakage, abnormal flow, insufficient air pressure and abnormal flame extinction in flame system
  • Protection function and alarms for insufficient carrier gas and protective gas pressure, insufficient cooling rate supply and over-heating in graphite furnace system
  • Adopting large-scale programmable logic array and Inter I2C bus technology
  • For long term reliability of the whole electronic system, provided with European type sockets and AMP adapters
  • AAS analysis is made under windows operating system which is easy-to-use with fast parameter setting and optimization
  • In order to increase the sensitivity in flame analysis, two high performance HCLs can be mounted on the lamp turret
  • The D2 lamp and S-H background correction capability at 1A and 1.8A respectively is better than 30 times
  • Provided with detector R928 photo multiplier with high sensitivity and wide spectral range
  • Data processing system: provided with analytical methods – Working curve auto-fitting standard addition method, automatic sensitivity correction, automatic calculation of concentration and content with windows operating system software
  • Experiments can be repeated 1~99 times along with automatic calculation of mean value, standard deviation and relative standard deviation
  • Equipped with multi task functions- sequential determination of multi-elements in the same sample, condition reading- with model function and result printing- measured data and final analytical report printout in excel
  • Provided with standard RS-232 serial port connection

Specification

O2 Enriched Flame Yes
Auto Sampler Yes
Flame Yes
High Performance Lamp Yes
Graphite Furnace Type Yes
Wavelength Range 190-900nm
Wavelength Accuracy Better than ±0.25nm
Resolution Two Spectral lines of Mn at 279.5nm and 279.8nm can be separated with the spectral bandwidth of 0.2nm and valley peak energy ratio less than 30%
Baseline Stability 0.004A/30min
Lamp Turret Motorized 6-lamp turret
Lamp Current Adjustment
Wide Pulse Current 0~25mA
Narrow Pulse Current 0~10mA
Lamp Power Supply Mode 400Hz Square wave pulse 100Hz Narrow Square wave pulse + 400Hz wide square wave pulse
Optical System
Monochromator Single beam, Czerny-Turner design grating Monochromator
Grating 1800l/mm
Focal Length 277mm
Blazed Wavelength 250nm
Spectral bandwidth 0.1nm, 0.2nm, 0.4nm, 1.2nm Auto Switch Over
Flame Atomizer
Burner 10cm single slot all-titanium burner
Spray Chamber Corrosion resistant all-plastic spray chamber
Nebulizer High efficiency glass nebulizer with metal sleeve, sucking up rate: 6-7ml/min
Emission Burner It is Provided
Graphite Furnace
Temperature Range RT~3000ºC
Heating Rate 2000ºC/sec
Graphite Tube Dimensions (L*OD) 28*8 mm
Characteristic Mass Cd≤0.8*10-12g, Cu≤5*10-12g, Mo≤1*10-11g
Precision Cd≤3%, Cu≤3%, Mo≤4%
Graphite Furnace Auto Sampler
Sample Tray Capacity 55 sample vessels and 5 reagent vessels
Vessel Material Polypropylene
Vessel Volume 3ml for sample vessel, 20ml for reagent vessel
Min. Sampling Volume 1µl
Repeatable Sampling Times 1-99 times
Sampling System Accurate dual pump system with 100µl and 1ml injectors
Air-C2H2 Flame Rich-Oxygen Air-C2H2 Flame
Cu: Characteristic Concentration ≤0.025mg/L
Cu: Detection Limit ≤0.006mg/L
Ba: Characteristic Concentration ≤0.22mg/L
Al: Characteristic Concentration ≤0.4mg/L
Main Unit Dimensions 1070 x 490 x 580 mm
Main Unit Weight 140kg
Graphite Furnace Dimensions 420 x 420 x 460 mm
Graphite Furnace Weight 65kg
Auto Sampler Dimensions 400 x 290 x 290 mm
Auto Sampler Weight 15kg